Spacelabs Healthcare Smart Disclosure System, Model 92810, a component of the Intesys Clinical Suite (ICS) G2.

Company
Spacelabs Healthcare, Inc.
Recall Initiated
January 28, 2019
Terminated
February 24, 2022
Recall Number
Z-2035-2019
Quantity
295 customer sites have the affected software versions
Firm Location
Snoqualmie, WA

Reason for Recall

Several reports were received that patient records were printed with the correct patient demographics but containing another patient's waveforms.

Distribution

Distribution was nationwide. There was also government and military distribution. Foreign distribution was made to Argentina, Bahrain, Bolivia, Brazil, Canada, Cayman Islands, Chile, China, Costa Rica, Czech Republic, France, Hong Kong, India, Indonesia, Italy, Kuwait, Malaysia, Mexico, Netherlands, Pakistan, Palestine, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, Spain, Switzerland, Taiwan, Trinidad and Tobago, Tunisia, Turkey, UAE, United Kingdom, and Viet Nam.

Lot / Code Info

All affected Smart Disclosure installations with software versions 5.00, 5.01, 5.02, or 5.03, UDI 10841522123580.

Root Cause

Software design

Action Taken

The recalling firm began issuing customer letters on 1/28/2019 via priority service, return receipt requested, to all U.S. customers. The letter was translated, as necessary, and emailed on 2/4/2019 to the international subsidiaries and distributors. The firm identified 6 new customers (5 U.S. and 1 international) and the letter for the 5 new U.S. customers was issued 5/30/2019 via priority service, return receipt requested. The letter for the 1 new international customer was emailed on 5/28/2019.