NORMFlow H-110 Irrigation Fluid Warmer, Continental Europe, Model H-1100, Reorder H-1100-INT-230. Product Usage: NORMOFLO H-1100 Irrigation Fluid Warmers has been designed for safe, in-line warming of irrigating fluids as they are administered to patients. The Fluid Warmer utilizes a solution reservoir housed in a controller unit (warmer). The recirculation solution is heated to present temperature in the reservoir, pumped though a heat exchanger ( a component of the Sets) and is returned to the reservoir for continuous recirculation. The NORMOFLO Warming products are typically used in clinical situations that may require rapid irrigation with warmed fluids.

Company
Smiths Medical Asd, Inc.
Recall Initiated
October 25, 2016
Posted
January 26, 2017
Terminated
July 25, 2017
Recall Number
Z-1105-2017
Quantity
1
Firm Location
Saint Paul, MN

Reason for Recall

Smiths Medical became aware that that one (1) NORMOFLO H-1100 Irrigation Fluid Warmers was incorrectly labelled with Spanish language labels instead of English labels. Specifically five (5) labels were incorrect; the Operators Manual, Outer Base Box label, Pole Assembly Box label, and two (2) O-ring Lube labels.

Distribution

Distribution in the foreign country of Sweden

Lot / Code Info

S105B0061

Root Cause

Labeling mix-ups

Action Taken

Consignee were hand delivered a Smiths Medical Urgent Medical device Filed Safety Notice letter dated October 25, 2016. The letter described the Reason for Field Corrective Action, Risk to Health, and Instructions to Customers. Advised consignees to Provide the device to the Smiths Medical Representative who delivered this notice to you. The Smiths Medical Representative will replace the labeling on the device and packaging, and provide you with the correct English language Operators Manual. Complete the Recall Confirmation Form and return it to Smith Medical via e-mail to FCA.Response@Smiths-medical.com. For questions regarding this notification, please ask the Smiths Medical Representative who delivered the notification or contact Smiths Medical Customer Service Department at +46 859477250.

More recalls by Smiths Medical Asd, Inc.

View all recalls by this company →