Container fill and transfer sets, 250 mL, 500 mL, 1000 mL, 2000 mL, 3000 mL, or 4000 mL.
This recall has been terminated (originally issued June 23, 2016).
- Company
- The Metrix Company
- Recall Initiated
- June 23, 2016
- Terminated
- April 11, 2023
- Recall Number
- Z-2260-2016
- Quantity
- 137,598 cases
- Firm Location
- Dubuque, IA
- Official Source
- View on FDA website ↗
Reason for Recall
An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.
Distribution
Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom
Lot / Code Info
All lots are being recalled. Lot code consists of the five digit product code followed by a dash followed by a unique five character numeric or alphanumeric sequence. Affected product codes for this product: 58781, 58785, 62005, 62010, 62015, 62025, 62030, 62035, 64005, 64010, 64015, 64025, 64030, 64035, 64781, 65005, 65010, 65015, 65025, 65030, 65035, 66005, 66010, 66015, 66020, 66025, 66030, 66035, 66305, 66310, 66315, 66325, 66330, 66335, 66705, 66710, 66715, 66725, 66730, 66735, 66910, 66915, 66925, 66930, 66935, 67005, 67010, 67015, 67025, 67030, 67035, 69005, 69010, 69015, 69025, 69030, 69035
Root Cause
Device Design
Action Taken
On 6/23/2016, the recalling firm sent written recall notices to their distributors and end users notifying them of the recall. Customers and consignees were instructed to not use the devices, locate and quarantine all affected devices, and to complete the response form.