RecallDepth

Junction Boxes, Product Usage: The Invacare Bariatric home care bed is intended to provide a rest surface with adjustable height from floor and adjustable surface contours.

Company
Invacare Corporation
Recall Initiated
March 15, 2016
Terminated
September 15, 2017
Recall Number
Z-2402-2016
Quantity
1,011 units
Firm Location
Elyria, OH

Reason for Recall

The firm recently identified a quality issue with the Junction box used in the Invacare Bariatric Bed. They discovered that the Junction Box may omit sparks which could result in a fire in the presence of flammable materials.

Distribution

US Nationwide Distribution including DC and PR. Internationally Canada.

Lot / Code Info

Model #'s:1104343 & 1104345

Root Cause

Nonconforming Material/Component

Action Taken

Invacare sent an Urgent Medical Device Recall letter dated May 12, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions call Invacares recall support line at (877) 413-6008, Monday Friday, 8 a.m. 5 p.m. EST, for assistance.

More recalls by Invacare Corporation