Junction Boxes, Product Usage: The Invacare Bariatric home care bed is intended to provide a rest surface with adjustable height from floor and adjustable surface contours.
- Company
- Invacare Corporation
- Recall Initiated
- March 15, 2016
- Terminated
- September 15, 2017
- Recall Number
- Z-2402-2016
- Quantity
- 1,011 units
- Firm Location
- Elyria, OH
Reason for Recall
The firm recently identified a quality issue with the Junction box used in the Invacare Bariatric Bed. They discovered that the Junction Box may omit sparks which could result in a fire in the presence of flammable materials.
Distribution
US Nationwide Distribution including DC and PR. Internationally Canada.
Lot / Code Info
Model #'s:1104343 & 1104345
Root Cause
Nonconforming Material/Component
Action Taken
Invacare sent an Urgent Medical Device Recall letter dated May 12, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions call Invacares recall support line at (877) 413-6008, Monday Friday, 8 a.m. 5 p.m. EST, for assistance.