Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal fixation device instrument.

Company
Rti Surgical, Inc. (Dba Pioneer Surgical Technology, Inc.)
Recall Initiated
April 22, 2016
Posted
May 26, 2016
Terminated
February 8, 2017
Recall Number
Z-1829-2016
Quantity
17 devices
Firm Location
Marquette, MI

Reason for Recall

Potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip, resulting in possible patient injury.

Distribution

US nationwide distribution in CA, FL, IN, MI, MO, NC, NY, and TX.

Lot / Code Info

Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve; Lot Number: 170684; Part Number: 26-COUNTER-T-XLINK

Root Cause

Process design

Action Taken

RTI sent an Urgent Voluntary Product Notification dated April 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. RTI Surgical is conducting a voluntary recall of the Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve due to the potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip resulting in possible patient injury. For further questions, please call (906) 226-4489.