Dialog+ Hemodialysis machines: Dialog A + HE/BIC Product Code 710200L Serial # >200318 and < 2039222 The Dialog+ can be used for implementing and monitoring hemodialysis treatments for patients with acute or chronic kidney failure. The system can be used in hospital, health center and outpatient dialysis center settings when prescribed by a Physician. The following types of renal replacement therapy can be carried out with the system: Hemodialysis (HD) with or without phases of pure ultrafiltration, High flux hemodialysis and low flux hemodialysis.

This recall has been terminated (originally issued May 3, 2016).

Company
B Braun Medical Inc
Recall Initiated
April 1, 2016
Posted
May 3, 2016
Terminated
July 11, 2017
Recall Number
Z-1531-2016
Quantity
370 units
Firm Location
Bethlehem, PA
Official Source
View on FDA website ↗

Reason for Recall

Potential leakage of conductivity sensors of the Dialog Dialysis machines.

Distribution

Nationwide Distribution to ND, OH, CA, VA, FL, LA, SC, IL, UT, AZ, KY, IA, WI, OK, TN, SD, AL, NJ, TX, NM, NY, KS, WA, MO, CO, PA, CT, DE, GA, MD, VA, Puerto Rico, and St. Thomas USVI

Lot / Code Info

Product Code 710200L, Serial Number Range 200318 -203922 distributed between 08/05/2013 - 07/07/2014.

Root Cause

Process control

Action Taken

B. Braun mailed an Urgent Medical Device Correction notice to affected customers informing them of the Dialog+Hemodialysis System due to the potential for Bicarbonate and End conductivity cells (3456102A and 3456103A respectively) to exhibit cracks. Notice provided customers with actions for Qualified Service Technicians, Clinical Personnel Administering Dialysis treatment and adverse event reporting. All Dialog+Hemodialysis machines will be inspected at the customer's site by the customer Biomed, who has been trained and qualified by B. Braun Avitum. The inspection of each machine will be performed in accordance to a Field Service Information (FSI). Customers with questions were instructed to contact the Technical Support Team at 800-621-0445 or the Customers Service Team at 800-848-2066. For questions regarding this recall call 610-997-4717.