Mizuho OSI Trios, Table Base; Voltage: 100-240V ~ 50/60 Hz Amps: 5,0-5.0A; Model 7803 and 7803AP Orthopaedic: Operating room tablebase - an electronically powered mobile operating table is created, designed for temporary (<24 hours) support and positioning of a patient in a prone, supine, or lateral position depending on the specific table top utilized.

Company
Mizuho Osi
Recall Initiated
December 14, 2015
Posted
February 9, 2016
Terminated
February 10, 2016
Recall Number
Z-0792-2016
Quantity
33
Firm Location
Union City, CA

Reason for Recall

Table Bases may have been assembled incorrectly, missing a retaining ring from the head end assembly. Over time, the parts could go out of alignment, making clutch engagement or disengagement difficult. If the clutch cannot disengage, the user would be unable to rotate the patient from supine to prone.

Distribution

US, South Africa, Japan and United Arab Emirates.

Lot / Code Info

Model numbers 7803, 7803AP.  Serial numbers 228-261.

Root Cause

Employee error

Action Taken

All consignees were notified by telephone and a recall letter was mailed to each facility as a follow up. Customers were asked to remove tables from operation until they were inspected by OSI service personnel.