RecallDepth

Copper NiTi Preformed Archwires; Copper NiTi 40 BAF .016X.022 LO Pk10, Part No. 211-0952. Intended to provide tooth movement.

This recall has been terminated (originally issued November 30, 2015).

Company
Ormco/Sybronendo
Recall Initiated
November 4, 2015
Posted
November 30, 2015
Terminated
January 19, 2016
Recall Number
Z-0343-2016
Quantity
161 units
Firm Location
Glendora, CA
Official Source
View on FDA website ↗

Reason for Recall

Ormco Corporation is recalling Copper NiTi Preformed Archwires because the product transformation temperature as listed on the labeling is incorrect.

Distribution

Distributed US (nationwide) and the countries of Canada, Japan, Mexico, Russia, and Peru.

Lot / Code Info

Lot No. 21538339

Root Cause

Employee error

Action Taken

A recall letter dated 11/5/15 was sent to all their customers who purchased the Copper NiTi Preformed Archwires because it has been brought to Ormco's attention that the product transformation temperature as listed on the labeling is incorrect. The wire is labeled as having 40 degrees Celsius transformation temperature, but the actual wire has a transformation temperature of 29 degrees Celsius. The letter provides the customers with the problems identified and the actions to be taken. Customers are instructed to complete and fax back the enclosed acknowledgement form.