RecallDepth

Viking M, L, and XL Mobile Lifts. Non-AC-powered patient lift.

This recall has been terminated (originally issued June 19, 2015).

Company
Hill-Rom, Inc.
Recall Initiated
March 20, 2015
Posted
June 19, 2015
Terminated
August 4, 2017
Recall Number
Z-1818-2015
Quantity
24, 798
Firm Location
Batesville, IN
Official Source
View on FDA website ↗

Reason for Recall

Complaints (including one reported death) allegedly of the lift arm drifting down suddenly. If the lift arm assembly is manually lifted, the actuator can become damaged and get stuck in the highest position. If a patient is lifted into the sling while the actuator is stuck, there is a potential risk of a free fall of the patient, resulting in minor or potentially catastrophic injuries.

Distribution

Worldwide Distribution -- AE, AO, AR, AT, AU, AZ, BD, BE, BG, BH, BR, CA, CH, CL, CN, CO, CS, CZ, DE, DK, EG, ES, FI, FR, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KR, KW, LB, LI, LT, LU, LV, MA, MT, MX, MY, NL, NO, NZ, OM, PH, PL, PT, QA, RO, SA, SE, SG, SI, SK, SY, TH, TN, TR, TW, UK, US, VE, and ZA.

Lot / Code Info

Viking¿ M Lift  Model 2040035 (S/N 9 200 000 - 9 201 689)  Model 2040015 (S/N 7 500 401 - 7 568 899)  Model 2040005 (S/N 7 100 101 - 7 200 200) Viking¿ L Lift  Model 2040004 (S/N 7 200 201 - 7 300 300) Viking¿ XL/Viking¿ 300 Lift  Model 2040003 (S/N 800 001 - 804 999)

Root Cause

Labeling design

Action Taken

On 3/20/2015, URGENT Field Safety Notice notifications, dated 3/19/2015 were sent to US customers by e-mail and or courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers are forward a copy of the letter to any other appropriate facility personnel. Additionally, the attached form should be completed and returned to Hill-Rom. Upon receipt, the firm will contact customers to schedule an inspection of the potentially affected devices. For questions, call 800-445-3720 (Mon - Thurs, 8 am - 6:30 pm; Friday, 8 am - 6 pm).

More recalls by Hill-Rom, Inc.

View all recalls by this company →