ORTHOFIX Firebird Spinal Fixation System 6.5mm x 55mm Multi-Axial Screw, Standard Implant Grade Titanium Alloy Rx Only Provided Non-Sterile Do Not Reuse

This recall has been terminated (originally issued June 10, 2015).

Company
Orthofix, Inc
Recall Initiated
May 18, 2015
Posted
June 10, 2015
Terminated
October 30, 2015
Recall Number
Z-1749-2015
Quantity
45 units
Firm Location
Lewisville, TX
Official Source
View on FDA website ↗

Reason for Recall

The Firebird Spinal Fixation System 6.5 mm x 55 mm Bone Screw, Self-Tapping is currently mislabeled at the distal tip of the bone screw showing a length measurement of 50 mm instead of 55 mm.

Distribution

US Distribution to the states of : OK, MD, CA, FL, MI, SC, CO, NY and PA., and Internationally to Asturias Spain.

Lot / Code Info

Lot Number S29

Root Cause

Unknown/Undetermined by firm

Action Taken

Orthofix sent an Urgent Medical Device Recall Notification letter dated May 15, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Affected customers were instructed to inspect their inventory, and identify and remove the recalled product to send back to Orthofix. The distributors are also instructed to provide recall notification to hospitals, surgeons, or other distributors that they have further distributed affected product to. Contact your local Orthofix representative or Orthofix customer service representative at 888-298-5700 for return instructions and shipment information.