DiaSorin, PRO-Trac II Tacrolimus ELISA kit, REF 32400 and 32400-CN. Product Usage: The PRO-Trac II Tacrolimus ELISA kit is an in vitro reagent system intended for the quantitative determination of tacrolimus (Prografl, FK506) and some metabolites in human EDT A or heparinized whole blood as an aid in the management of liver transplant patients receiving tacrolimus therapy.

Company
Diasorin Inc.
Recall Initiated
March 16, 2015
Posted
April 20, 2015
Terminated
September 15, 2015
Recall Number
Z-1464-2015
Quantity
451 kits
Firm Location
Stillwater, MN

Reason for Recall

Instability in the PRO-Trac II Tacrolimus ELISA kit which may cause falsely elevated patient results.

Distribution

Worldwide Distribution - US in the state of CA and the countries of: CHINA, BELGIUM, ARGENTINA, and INDIA.

Lot / Code Info

131716, 131716A, 131716B, 131716C, 131716D, 131716E,  131717.

Root Cause

Under Investigation by firm

Action Taken

DiaSorin sent a Customer Notification Letter" to affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were asked to complete and return the confirmation form to the manufacturer. For questions contact DiaSorin Inc. Product Support at 1-800-328-1482.