RecallDepth

Terumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375 Product Usage: Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4: I ratio. Includes a Conducer Heat Exchanger and a MP4 Temperature & Pressure Module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity.

This recall has been terminated (originally issued October 31, 2014).

Company
Terumo Cardiovascular Systems Corp
Recall Initiated
March 1, 2011
Posted
October 31, 2014
Terminated
November 6, 2014
Recall Number
Z-0160-2015
Quantity
Catalog: 16579 ; Catalog 6375: 3256 units
Firm Location
Ashland, MA
Official Source
View on FDA website ↗

Reason for Recall

Defect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line

Distribution

Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.

Lot / Code Info

Catalog Number: 165720 Lot Numbers: MDI5, MD15B, MD22, ME19, MP13. Manufacturing dates March 15, 20 I 0 through December 13, 20 I 0.  Catalog Number: 6375 Lot Numbers: MD15, ME19, MPI3  Manufacturing dates March 15, 2010 through December 13, 20 I 0

Root Cause

Device Design

Action Taken

Terumo Cardiovascualr notified accounts by letter dated 3/1/2011 advising users users that the luer connector to the table line in certain lots of myocardial protection sets may leak. Users should not continue to use affected myocardial protection sets in a manner that exposes the connector to negative pressure without taking correction or replacement activities as described. Firm provided mitigating instructions: Clamp the table line distal to the luer connector when ending an infusion of cardioplegia thus isolating the leak from negative pressure. Contact Terumo CVS Customer Service to report the defect and arrange return of the unit: 800-521-2818.