RENASYS EZ Negative Pressure Wound Therapy (NPWT) System (RENASYS EZ) for patients who would benefit from a suction device particulary as the device may promote wound healing.

Company
Smith & Nephew Inc.
Recall Initiated
August 13, 2014
Posted
September 23, 2014
Terminated
August 6, 2015
Recall Number
Z-2688-2014
Quantity
2,142 units
Firm Location
Saint Petersburg, FL

Reason for Recall

Modification were made to the RENASYS EZ NPWT Systems without a 510(k) or PreMarket Approval notification.

Distribution

Worldwide Distribution- US (nationwide) including MA, MN, NY and Puerto Rico; Canada, Mexico, Dubai, Australia, Chile, Colombia, Germany, France, England, Spain, Italy, Norway, Hong Kong, Shanghai, New Zealand, South Africa, Singapore, Japan, Thailand and the Caribbean Islands.

Lot / Code Info

Model #'s 66800059, all lots since product launched 01/16/2009

Root Cause

No Marketing Application

Action Taken

All RENASYS EZ Products will be removed from the field and customers will be transitioned to other product. On August 13, 2014 the firm sent an "Urgent- Medical Device Removal RENASYS EZ" notification via FedEx overnight asking that the use of the RENASYS EZ should be immediately discontinued and the device should be returned to Smith & Nephew. If you have questions, please reach out to your Customer Care representative or call 1-800-876-1261 Monday through Friday, 8:00 AM to 6:00 PM, Eastern Time.