ProVue Instrument Driver for JResultNet (ProVue.jrm) v. 1.5.3 and below JResultNet is a stand-alone software package that provides connectivity between in-vitro diagnostic instruments (IVDs) and Laboratory Information Systems (LIS) aiding the flow of laboratory information, including test orders and results, specimen information, and patient information.

This recall has been terminated (originally issued September 12, 2014).

Company
Data Innovations, Llc
Recall Initiated
August 15, 2014
Posted
September 12, 2014
Terminated
March 4, 2015
Recall Number
Z-2637-2014
Quantity
84
Firm Location
Fort Myers, FL
Official Source
View on FDA website ↗

Reason for Recall

Potential issues with the ProVue driver have been identified which may cause data to be misassociated with patient results under a given set of circumstances.

Distribution

Nationwide Distribution including AL, AZ, CA, CT, DC, FL, GA, IN, LA, MI, MO, NC, NH, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, and WA.

Lot / Code Info

Version #s: 1.0.0 Date Released 6/17/2004, 1.0.1 Date Released 8/18/2004, 1.1.1 Date Released 9/9/2004, 1.2.0 Date Released 10/11/2004, 1.3.0 Date Released 11/1/2004, 1.3.1 Date Released 1/20/2005, 1.3.2 Date Released 2/18/2005, 1.4.0 Date Released 3/30/2005, 1.4.1 Date Released 8/31/2005, 1.4.2 Date Released 9/12/2005, 1.4.2 Date Released 10/5/2005, 1.4.3 Date Released 2/8/2006, 1.4.4 Date Released 2/20/2006, 1.4.5 Date Released 3/3/2006, 1.4.5 Date Released 3/23/2006, 1.4.5 Date Released 5/17/2006, 1.5.0 Date Released 6/26/2007, 1.5.1 Date Released 9/17/2007, 1.5.1 Date Released 12/13/2007, 1.5.1 Date Released 11/7/2008, 1.5.2 Date Released 4/13/2009, 1.5.3 Date Released 1/12/2011.

Root Cause

Software design

Action Taken

Data Innovations LLC sent an Important Medical Device Information letter on August 15, 2014, to consignees for the JResultNet using ProVue Instrument Driver JResultNet (provue.jrm) v. 1.5.3 and below. Potential issues with the ProVue driver have been identified which may cause data to be mis-associated with patient results under a given set of circumstances. The user on the ProVue transmits the same results twice back to back by clicking the Send to Host button twice. The ProVue sends two ASTM messages back to back before ending transmission (sending an EOT character) The solution was to disconnect JResultNet from the ProVue instrument and apply the patch to driver provue.jrm when it is available. If you have further questions regarding this notice, please feel free to contact us at northamerica-support@datainnovations.com or 239-931-6004 Toll-Free (from the US and Canada): 1-800-332-0499.