INSIGHT Dental Film, Catalog Number/REF 811 0785, KODAK and Carestream DENTAL labels --- Carestream Health Inc.

Company
Carestream Health, Inc.
Recall Initiated
July 2, 2014
Posted
August 1, 2014
Terminated
May 25, 2016
Recall Number
Z-2142-2014
Quantity
1633 boxes
Firm Location
Rochester, NY

Reason for Recall

The Indicator Dot on the film packet's outer printed paper labeling was incorrectly positioned relative to the location of the raised dot on the dental intraoral film. This may cause confusion for the dental practitioner. If the processed film is placed on the view box by referencing the film's raised dot, the image orientation will be rotated 180 degrees.

Distribution

Worldwide Distribution.

Lot / Code Info

Lot Numbers: 53901201, 53801303, 53801303-1, 54601201, 54601201-1

Root Cause

Package design/selection

Action Taken

Recall notices were mailed to US Customers via overnight carrier on 07/02/2014. Customers are asked to complete an inventory of goods to be returned which will be ultimately destroyed by recalling company. If you have any questions or concerns, please contact the Carestream Customer Care Center in the U.S. at 1-800-933-8031.