RecallDepth

RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 799, 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.

Company
Med-Mizer, Inc.
Recall Initiated
April 18, 2014
Posted
May 2, 2014
Terminated
May 23, 2016
Recall Number
Z-1542-2014
Quantity
1
Firm Location
Batesville, IN

Reason for Recall

All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on th

Distribution

USA (nationwide)

Lot / Code Info

Serial Number 2000

Root Cause

Process change control

Action Taken

On 5/1/14, Med-Mizer sent URGENT: MEDICAL DEVICE RECALL notification to their consignees. Notification included description of the affected product, reason for recall, risk to health, instructions on how to recognize how the device may fail, and instructions for actions to be completed by their consignees. Contact Information: 1-877-867-7365, sales@med-mizer.com, 8am-5pm EST.

More recalls by Med-Mizer, Inc.

View all recalls by this company →