RecallDepth

Optiflux B200 Flux Dialyzers; Product Number: 0500325E. Single Use acute or chronic hemanalysis treatment.

This recall has been terminated (originally issued February 26, 2014).

Company
Fresenius Medical Care Holdings, Inc.
Recall Initiated
February 4, 2014
Posted
February 26, 2014
Terminated
May 18, 2021
Recall Number
Z-1105-2014
Quantity
100,008 units
Firm Location
Waltham, MA
Official Source
View on FDA website ↗

Reason for Recall

Dialysate Port Leak During Priming of the Dialyzers

Distribution

Worldwide distribution: USA (nationwide) and the country of: Mexico.

Lot / Code Info

Lot Numbers: 13LU02011 13LU02013 13NU02016 13NU02017

Root Cause

Under Investigation by firm

Action Taken

Fresenius Medical issued Notice to Clinics and Clinics with Patients Letters Dated - 1st Mailing - February 4, 2014 and Patients Letters Dated - February 5, 2014 - these letters contained the correct information.Corrected letters to Clinics and Clinics with Patients issued 2/14/14 via certified mail with return receipt and fax-back form . Users are advised the leak may be stopped by tightening the dialysate port cap more securely. Questions associated with this product notification to your Customer Service Team at 1-800-323-5188.

More recalls by Fresenius Medical Care Holdings, Inc.

View all recalls by this company →