Morph AccessPro Steerable Introducer; 6F Introducer, 45 cm working length Single use only. BioCardia, Inc. San Carlos, CA 94070 BioCardia Morph AccessPro is intended to provide a pathway through medical instruments cush as balloon dilation catheters, guidewires or other therapeutic devices that may be introduced into the peripheral vasculature.

This recall has been terminated (originally issued March 27, 2014).

Company
Biocardia, Inc.
Recall Initiated
March 3, 2014
Posted
March 27, 2014
Terminated
June 19, 2014
Recall Number
Z-1301-2014
Quantity
~2541 - both models
Firm Location
San Carlos, CA
Official Source
View on FDA website ↗

Reason for Recall

Firm discovered the potential for particulates in the inner lumen of Morph Access Pro Steerable Introducer, which could result in am embolic event.

Distribution

Worldwide Distribution: US (nationwide) and country of: Israel.

Lot / Code Info

Catalog number MAP645 -  Lot numbers:  1037  1048  1055  1059  1069  1073  1074  1079  1092  1095  1115  1118  1121  1125  1128  1130  1134  1147  1155  1164  1168  1169  1173  1177  1184  1190

Root Cause

Other

Action Taken

The firm, BioCardia, Inc., sent an "URGENT-VOLUNTARY MEDICAL DEVICE RECALL" letter dated March 3, 2014 its customers, hospital's Director of Risk Management, via USPS mail and priority mail to Cardiac Catheterization Laboratory. The letter described the product, problem and actions to be taken. The customers were instructed to immediately examine your inventory; quarantine the product; return the product to BioCardia and contact Customer Service at 650-226-0148 to coordinate product return; in the event you no longer have this device as a result of further distribution, identify the purchaser and notify them at once of this recall; and complete and return the enclosed MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement and Receipt Form via FAX to 650-631-3731 or mail to: BioCardia, Inc., 125 Shoreway Road, Suite B, San Carlos, CA 94070 as soon as possible. If you have any questions, call the Vice President, Quality Assurance at 650-226-0143.