PrepStain Tecan US-II. For use in the screening and detection of cervical cancer.

This recall has been terminated (originally issued August 28, 2013).

Company
Becton Dickinson & Co. Bd Diagnostic Systems
Recall Initiated
July 8, 2013
Posted
August 28, 2013
Terminated
May 1, 2014
Recall Number
Z-2112-2013
Quantity
890 instruments
Firm Location
Sparks, MD
Official Source
View on FDA website ↗

Reason for Recall

PrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative.

Distribution

Worldwide Distribution-USA (nationwide) and Puerto Rico and the country of Canada.

Lot / Code Info

Catalog Numbers - 799-14000-00 (490100) and 799-14000-00R (490407)

Root Cause

Software design

Action Taken

Ex-US BD sites were notified via email and teleconference, and will communicate directly with customer sites in their region as required by local regulation. All US customers were initially contacted via telephone and letters were mailed.