HemosIL RecombiPlasTin 2G, Part Number 0020002950 (8 ml size). The product is used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Anticoagulant Therapy.

This recall has been terminated (originally issued July 22, 2013).

Company
Instrumentation Laboratory Co.
Recall Initiated
April 9, 2013
Posted
July 22, 2013
Terminated
May 6, 2015
Recall Number
Z-1784-2013
Quantity
Kits per product:1)11080, 2)11603, 3)11487, 4)11483, 5)11290, 6)11562, 7)11535, 8)11577, 9)11505, 10)11708,11)11579,12)11580, 13&14) not distributed
Firm Location
Bedford, MA
Official Source
View on FDA website ↗

Reason for Recall

Instrumentation Laboratory is recalling HemosIL RecombiPlasTin 2G (RPT2G), Part No. 0020009580 (8 mL Size) due to being Out of Specification.

Distribution

Worldwide Distribution- US (nationwide) including states of: MA, ME, CT, NJ, NY, PA, MD, VA, NC, SC, GA, FL, AL, TN, MS, KY, OH, IN, MI, IA, WI, MN, SD, MT, IL, MO, KS, NE, LA, AR, OK, TX, CO, ID, AZ, CA, OR, WA, AK, RI, VT and the countries of: Angola, Argentina, Australia, Austria, Belgium, Bermuda, Brazil, Bulgaria, Canada, Caribbean, Chile, Colombia, Costa Rica, Czech, Denmark, Dubai, Ecuador, Egypt, France, Germany, Gibraltar, Great Britain, Greece, Guatemala, Hungary, India, Indonesia, Iran, Israel, Italy, Jamaica, Jordan, Korea, Kosovo, Kuwait, Lebanon, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, Norway, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad, Turkey, Ukraine, Uruguay, Vietnam, and Japan.

Lot / Code Info

Part Number applies to all: 0020002950 - all product lots.  Lot Numbers: 1) N0716706 Exp. Date 5/31/2013, 2) N0817321 Exp. Date 5/31/2013, 3) E0917396 Exp. Date 9/30/2013, 4)) N0917378 Exp. Date 9/30/2013, 5) N1118029 Exp. Date 10/31/2013, 6) N0128669 Exp. Date 1/31/2014, 7) N0329291 Exp. Date 2/28/2014, 8) N0429596 Exp. Date 3/31/2014, 9) N0520056 Exp. Date 5/31/2014,10) E0720822 Exp. Date 7/31/2014, 11) N0821174 Exp. Date 8/31/2014, 12) N091388 Exp. Date 9/30/2014, 13) N0921522 Exp. Date 9/30/2014, 14) N0132676 Exp. Date 1/31/2015 and 15) N1222567 Exp. Date 1/31/2015.

Root Cause

Nonconforming Material/Component

Action Taken

Between April 9, 2013 and April 11, 2013, Instrumentation Laboratory distributed "Important Product Correction" notification letters to their U.S. and foreign customers informing them that the on-board instrument stability claim has been reduced to 3 days for all current and future product lots of HemosIL RecombiPlasTin 2G. A second notification to non-respondents was initiated and completed on May 7, 2013. Technical questions were referred to 1-800-678-0710, Option # 2.