BS-200 Chemistry Analyzer Designed for clinical laboratory use.

Company
Mindray Ds Usa, Inc. D.B.A. Mindray North America
Recall Initiated
November 28, 2012
Posted
March 22, 2013
Terminated
June 12, 2013
Recall Number
Z-0977-2013
Quantity
3 units
Firm Location
Mahwah, NJ

Reason for Recall

An issue has been identified with the BS-200 where clinical chemistry results may be inaccurate.

Distribution

Distributed in the states of NC and TX.

Lot / Code Info

Part number BA20E-PA00003; Serial Numbers: WN-28103978F, WN-28103976F, and WN-28103977F.

Root Cause

Device Design

Action Taken

Mindray North America /Mindray DS USA Inc. sent Urgent Product Corrective Action letters via certified mail, return receipt requested, signature required on 11/28/2012. The letters identified the affected product as well as the issue, effect on patients, and the corrective action. Customers are to remove the analyzer from service until the affected pump has been replaced. Also, customers are to contact Mindray Customer Service Representative Tony Li at 1-800-288-2121 x2112 for technical assistance, and to arrange for replacement pumps at no charge.