BS-200 Chemistry Analyzer Designed for clinical laboratory use.
- Recall Initiated
- November 28, 2012
- Posted
- March 22, 2013
- Terminated
- June 12, 2013
- Recall Number
- Z-0977-2013
- Quantity
- 3 units
- Firm Location
- Mahwah, NJ
Reason for Recall
An issue has been identified with the BS-200 where clinical chemistry results may be inaccurate.
Distribution
Distributed in the states of NC and TX.
Lot / Code Info
Part number BA20E-PA00003; Serial Numbers: WN-28103978F, WN-28103976F, and WN-28103977F.
Root Cause
Device Design
Action Taken
Mindray North America /Mindray DS USA Inc. sent Urgent Product Corrective Action letters via certified mail, return receipt requested, signature required on 11/28/2012. The letters identified the affected product as well as the issue, effect on patients, and the corrective action. Customers are to remove the analyzer from service until the affected pump has been replaced. Also, customers are to contact Mindray Customer Service Representative Tony Li at 1-800-288-2121 x2112 for technical assistance, and to arrange for replacement pumps at no charge.