Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels.

This recall has been terminated (originally issued February 28, 2013).

Company
Orthofix, Inc
Recall Initiated
February 1, 2013
Posted
February 28, 2013
Terminated
July 1, 2013
Recall Number
Z-0896-2013
Quantity
565 total (287 PN 52-6800 and 278 PN 52- 6700)
Firm Location
Lewisville, TX
Official Source
View on FDA website ↗

Reason for Recall

There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.

Distribution

Worldwidewide Distribution - USA Nationwide including Puerto Rico and the countries of: Australia, Lebanon, Germany, Spain, Sweden and Israel.

Lot / Code Info

Catalog Numbers: 52-6800 & 52-6700; All Lot Numbers

Root Cause

Device Design

Action Taken

Orthofix sent an Urgent Medical Device Recall Notification dated February 1, 2013 to all affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to identify and remove the affected products from inventory immediately and follow the actions listed in the enclosed Product Return Instructions. For questions or concerns contact Regulatory Affairs at 214-937-2100.