RoboCouch Patient Positioning System Product Usage: is intended for use in the support and positioning of a patient during radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required.

This recall has been terminated (originally issued April 20, 2013).

Company
Accuray Incorporated
Recall Initiated
March 27, 2013
Posted
April 20, 2013
Terminated
August 20, 2013
Recall Number
Z-1155-2013
Quantity
16 devices are subject to correction.
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

A5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) motions of the RoboCouch manipulator may be defective.

Distribution

Worldwide Distribution - USA Nationwide including the states of: OH, MO. OK, CA and VA and the countries of: Finland, France, Luxembourg, Poland, Russia, Turkey and Japan.

Lot / Code Info

Model number: 025007

Root Cause

Nonconforming Material/Component

Action Taken

Accuray issued an Urgent Advisory Notification letter dated March 27, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that Field Service Engineers will be replacing the affected product. Questions and concerns are to be directed to Accuray Customer Support at 1-877-668-8667