Carestream DRX-Revolution Mobile X-Ray System -- Made in U.S.A. by Carestream Health Inc. 150 Verona Street, Rochester, NY . 14608 --- The DRX Revolution Mobile system is a mobile diagnostic x-ray system for generation of x-rays for examination of various anatomical regions. These products are mobile diagnostic x-ray systems designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas.

Company
Carestream Health, Inc.
Recall Initiated
August 29, 2013
Posted
January 3, 2014
Terminated
November 18, 2015
Recall Number
Z-0491-2014
Quantity
310 systerms domestically
Firm Location
Rochester, NY

Reason for Recall

It was discovered that Carestream received a report from a customer that 2 different patients were re-exposed due to the fact that the processed images were not able to be found in the system.

Distribution

Nationwide

Lot / Code Info

DRX Revolution Mobile X-Ray Systems manufactured from June 2012 through October 2013: Serial Numbers DRXR00690, 111, 114, 119, 121, 122, 123, 125, 126, 134, 142, 143, 151, 152, 153, 154, 155, 161, 162, 163, 164, 165, 166, 167, 168, 169, 173, 174, 176, 177, 178, 180, 181, 182, 183, 188, 189, 190, 191, 192, 196, 197, 198, 199, 201, 202, 203, 205, 206, 207, 208, 209, 211, 221, 222, 223, 224, 225, 226, 227, 228, 229, 230, 231, 232, 235, 236, 237, 238, 239, 240, 241, 242, 243, 244, 245, 246, 247, 248, 249, 254, 261, 262, 263, 264, 265, 266, 267, 268, 269, 270, 275, 276, 279, 280, 281, 282, 283, 284, 285, 286, 305, 312, 316, 317, 318, 319, 320, 321, 322, 323, 324, 325, 326, 330, 331, 332, 333, 334, 339, 340, 341, 342, 343, 344, 345, 346, 347, 348, 349, 353, 355, 357, 358, 359, 360, 361, 362, 363, 364, 365, 372, 373, 375, 376, 379, 380, 381, 382, 383, 384, 385, 386, 388, 394, 395, 396, 397, 398, 399, 400, 401, 403, 405, 412, 413, 414, 415, 417, 420, 421, 423, 431, 432, 433, 434, 435, 441, 442, 443, 444, 451, 452, 454, 455, 456, 457, 458, 459, 460, 461, 464, 466, 467, 487, 488, 489, 490, 495, 506, 510, 511, 512, 514, 515, 516, 517, 518, 519, 520, 521, 522, 523, 524, 530, 538, 541, 542, 543, 544, 545, 546, 547, 548, 549, 555, 557, 560, 561, 562, 563, 564, 581, 583, 584, 590, 591, 592, 593, 594, 595, 596, 597, 598, 599, 601, 602, 603, 609, 610, 611, 612, 613, 614, 616, 619, 620, 621, 622, 626, 627, 628, 629, 631, 632, 633, 634, 635, 636, 637, 639, 644, 645, 649, 650, 651, 656, 657, 658, 659, 660, 661, 663, 664, 665, 676, 677, 682, 685, 686, 687, 688, 689, 691, 692, 694, 696, 697, 698, 699, 700, 702, 703, 704, 705, 706, 707, 718, 723, 724.

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Carestream sent a letter to all affected customers informing them of the issue, how to detect the issue and actions to be taken by Carestream to correct the issue. The actions to be taken by Carestreamn to correct the issue will include the installation of a software patch to mitigate the effects of the observed failure and remove the possibility of a re-exposure in this failure mode. The correction will be implemented within the US to all affected devices at identified customer sites. For further questions call Carestream Customer Care at 1-800-328-2910, 8 am through 8 pm Eastern Standard Time.