3.7mm and 5.0mm Dynamic Locking Screw (DLS) Product Usage: 3.7mm and 5.0mm Dynamic Locking Screw (DLS) in combination with Synthes Locking Compression Plates (LCP) are intended for use in long bone fractures, the fixation of osteopenic bone, the fixation of osteotomies, and for the fixation of non-unions and malunions.

Company
Synthes Usa (Hq), Inc.
Recall Initiated
July 22, 2013
Posted
November 26, 2013
Terminated
August 18, 2016
Recall Number
Z-0390-2014
Quantity
40248
Firm Location
West Chester, PA

Reason for Recall

All lots of 3.7mm and 5.0mm Dynamic Locking Screw (DLS) were recalled due to complaints of breakages at the distal tip of the pin identified after successful healing and during planned removal of the device.

Distribution

Worldwide Distribution - USA (nationwide) and internationally to Canada.

Lot / Code Info

All lots of parts 09.213.022S - 09.213.070S, and 09.223.032S- 09.223.090S.

Root Cause

Labeling False and Misleading

Action Taken

Synthes sent an Urgent Notice: Medical Device Recall letter to Sales Reps and End Users on July 22, 2013. The letter identified the affect product, problem and actions to be taken. For questions call the Recall center at 610-719-5450.