Fresenius Liberty Cycler Part Number: RTLR180111 Product Usage: The Liberty cycler is a portable device designed to perform acute and chronic peritoneal dialysis.

This recall has been terminated (originally issued November 5, 2013).

Company
Fresenius Medical Care Holdings, Inc.
Recall Initiated
October 14, 2013
Posted
November 5, 2013
Terminated
May 18, 2021
Recall Number
Z-0148-2014
Quantity
16,600
Firm Location
Waltham, MA
Official Source
View on FDA website ↗

Reason for Recall

Fluid leaking may be observed inside the pump compartment of the Liberty Cycler and may result in dialysate becoming contaminated, potentially causing peritonitis

Distribution

USA Nationwide Distribution

Lot / Code Info

All serial numbers

Root Cause

Device Design

Action Taken

Fresenius Medical Care contacted all affected consignees by telephone on 10/14/13 to inform them of the product safety alert. Follow-up Urgent Safety Alert letter was sent on 10/18/13 and 10/23/13 by Certified Mail, Return Receipt requested. The letter identified the affected product, problem and actions to be taken. Customers were instructed to report any events associated with this safety alert to Customer Service Team at 1-800-323-5188.