Plum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 11971; b) list 11973 - software version 10.3; c) list 12391 - software version 11.3; d) list 20679 - Hospira MedNet Software e) list 20792 - Hospira MedNet Software; Product Usage: The Plum A+ Infusion System is a dual-line volumetric infusion pump that provides precise delivery of multiple i.v. therapies across the general spectrum of clinical care, including ICU/CCU, OR, PACU, ER, Med/Surg, Pediatrics, Outpatient clinics and Home care

This recall has been terminated (originally issued December 31, 2012).

Company
Hospira Inc.
Recall Initiated
December 7, 2012
Posted
December 31, 2012
Terminated
November 14, 2016
Recall Number
Z-0624-2013
Quantity
297,833 units
Firm Location
Lake Forest, IL
Official Source
View on FDA website ↗

Reason for Recall

The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.

Distribution

Worldwide Distribution -- USA, including Puerto Rico, and the countries of Austria, Australia, Argentina, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, Finland, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates, and Vietnam.

Lot / Code Info

list numbers 11971, 11973, 12391, 20679, and 20792; all serial numbers

Root Cause

Device Design

Action Taken

Hospira Inc. sent "URGENT DEVICE FIELD CORRECTION" letters, dated December 7, 2012, to all affected customers. The letters identified the products, problem and action(s) to be taken until doors have been replaced. The customers were instructed to read the information provided carefully, including instructions to inspect the door roller assembly prior to loading a cassette for signs of damage. Customers are to complete the attached Reply Form and return it to the firm via fax at 1-877-650-8359. Customers were instructed to notify their consignees if they have further distributed the affected product. If customers require additional assistance, they are to contact Stericycle at 1-877-650-0362..