Elecsys Troponin I STAT Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human serum and plasma. Elecsys Troponin I is intended to aid in the diagnosis of myocardial infarction.

This recall has been terminated (originally issued April 11, 2012).

Company
Roche Diagnostics Operations, Inc.
Recall Initiated
March 12, 2012
Posted
April 11, 2012
Terminated
February 6, 2013
Recall Number
Z-1388-2012
Quantity
1,288 kits
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

An investigation has confirmed a considerably reduced recovery of Li-heparin plasma samples for Troponin I lot 163176 and Troponin I STAT lot 163177. In the worst case, Li-Heparin sample recovery may be as low as 50% of the serum recovery. Correct serum recovery has been confirmed for the affected lots. The issue is caused by the raw material lot (Poly-L-Lysin) instability. Poly-L-Lysin is the

Distribution

Worldwide distribution. Argentina, Kenya, Austria, South Korea, Belgium, Kuwait, Bahrain, Morocco, Brazil, Mexico, Switzerland, Nigeria, Chile, New Zealand, Columbia, Philippines, Germany, Pakistan, Algeria, Poland, Egypt, Qatar, Spain, Tunisia, France, Turkey, United Kingdom, Taiwan, India, Ukraine, Iran, Venezuela, Italy, South Africa, Jordan, and Armenia.

Lot / Code Info

Elecsys Troponin I STAT Immunoassay, 05094798160- lot 163177

Root Cause

Nonconforming Material/Component

Action Taken

The firm, Roche Diagnostics Corporation, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated March 12, 2012 to all customers that received the affected lots of Elecsys Troponin I or Elecsys Troponin I STAT. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of the affected product; discard the affected product from their inventory according to their site's local regulations; if their facility has distributed the affected product to other sites, ensure that this letter is provided to those sites; complete and return the attached fax form via fax to 1-888-912-8457 and file this letter for future references. If you have any questions about the information contained in this letter, please contact Roche Diagnostics Technical Support, 24 hours a day, seven days a week at 1-800-428-2336.