Abbott Vascular COPILOT Kit, COPILOT Bleedback Control Valve Part Number : 1003330. For use in cardiac diagnostic or therapeutic procedures.

Company
Abbott Vascular
Recall Initiated
January 7, 2011
Posted
July 12, 2011
Terminated
June 26, 2012
Recall Number
Z-2792-2011
Quantity
389 units total
Firm Location
Temecula, CA

Reason for Recall

Five lots of Guide Catheters and Accessory products were processed through Ethylene Oxide (ETO) sterilization with additional plastic material covering the pallet, which is not consistent with validated packaging configuration for ETO sterilization.

Distribution

Nationwide distribution, and Ireland.

Lot / Code Info

Lot Number: 0121191

Root Cause

Process control

Action Taken

Abbott Vascular sent an Urgent Device Recall letter, dated January 7, 2011, noting the products being recalled and the reasons for recall. The firm also provided the following reconciliation instructions to its territory manager: -Help customer identify aflected product. - CaIl Abbott Vascular Customer Service (800) 221-9902 to receive RGA number. Record RGA number below. - Fax this completed form to (951) 914-3826. - Return a copy of this completed form with the returned product