klinoPORT, AC powered Operating Table
This recall has been terminated (originally issued December 22, 2010).
- Recall Initiated
- October 26, 2010
- Posted
- December 22, 2010
- Terminated
- March 20, 2012
- Recall Number
- Z-0802-2011
- Quantity
- 15 units
- Official Source
- View on FDA website ↗
Reason for Recall
The motorized adjustable pendant arm may suddenly descend to its lowest position.
Distribution
Nationwide Distribution -- CT, MA, and MO.
Lot / Code Info
Model number: klinoPORT 706, 806, 906, 1006 and 1306, Serial number between: 101039290 and 101158669
Root Cause
Other
Action Taken
Consignees were notified by a Warning Notice on/about 10/26/2010. Medical facilities were instructed to adjust the pendant arm to the lowest height position and disable the motor and to place the accompanying warning label on the pendant. They will, therefore, not be able to adjust the height of the pendant until corrective action is developed and completed. TRUMPF Medical Systems, Inc. will receive corrective action repair kits by 11/12/2010.