klinoPORT, AC powered Operating Table

This recall has been terminated (originally issued December 22, 2010).

Company
Kreuzer Gmbh & Co. Kg Benzstrabe 26 Puchheim Germany
Recall Initiated
October 26, 2010
Posted
December 22, 2010
Terminated
March 20, 2012
Recall Number
Z-0802-2011
Quantity
15 units
Official Source
View on FDA website ↗

Reason for Recall

The motorized adjustable pendant arm may suddenly descend to its lowest position.

Distribution

Nationwide Distribution -- CT, MA, and MO.

Lot / Code Info

Model number: klinoPORT 706, 806, 906, 1006 and 1306, Serial number between: 101039290 and 101158669

Root Cause

Other

Action Taken

Consignees were notified by a Warning Notice on/about 10/26/2010. Medical facilities were instructed to adjust the pendant arm to the lowest height position and disable the motor and to place the accompanying warning label on the pendant. They will, therefore, not be able to adjust the height of the pendant until corrective action is developed and completed. TRUMPF Medical Systems, Inc. will receive corrective action repair kits by 11/12/2010.