TrueBeam and True Beam STx versions 1.0 through 1.5, Model number H19; Varian Medical Systems, Palo, Alto, CA 94304. TrueBeam Radiotherapy Delivery System is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

This recall has been terminated (originally issued March 6, 2012).

Company
Varian Medical Systems, Inc. Oncology Systems
Recall Initiated
January 30, 2012
Posted
March 6, 2012
Terminated
April 16, 2014
Recall Number
Z-1175-2012
Quantity
182 units ***12-12-12 AMENDED TO: 200 units***
Firm Location
Palo Alto, CA
Official Source
View on FDA website ↗

Reason for Recall

An anomaly has been identified with the respiratory gating software of TrueBeam. When importing breath-hold gating protocols, the gating thresholds can be reset to default values, rather than retaining the thresholds established during planning.

Distribution

Nationwide Distribution (USA) and the countries of Australia, Belgium, Canada, China, Denmark, Finland, France, Germany, Hong Kong, India, Israel, Italy, Netherlands, New Zealand, Norway, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Thailand, Turkey and the United Kingdom.

Lot / Code Info

Model H19: Code numbers:  H191182,H191035,H191150,H191100,H191103,H191040,H191072,H191065, H191089,H191116,H191068,H191160,H191141,H191113,H191027,H191145, H191026,H191093,H191054,H191189,H191129,H191090,H191197,H191022,H191063,H191010,H191162,H191078,H191098,H191142,H191066,H191005,H191018,H191076,H191188,H191048,H191137,H191064,H191153,H191044,H191151,H191046,H191136,H191067,H191031,H191077,H191158,H191028,H191156,H191088,H191132,H191009,H191069,H191085,H191186,H191101,H191094,H191039,H191124,H191073,H191183,H191008,H191133,H191016,H191003,H191071,H191175,H191152,H191168,H191099,H191166,H191029,H191036,H191155,H191092,H191034,H191043,H191045,H191074,H191170,H191130,H191055,H191147,H191178,H191114,H191107,H191164,H191047,H191015,H191091,H191019,H191038,H191163,H191203,H191110,H191061, H191049,H191179,H191096,H191079,H191106,H191017,H191154,H191006,H191024,H191108,H191149,H191052,H191199,H191023,H191117,H191033,H191104,H191131,H191011,H191118,H191087,H191109,H191053,H191187,H191126,H191191,H191042,H191080,H191161,H191105,H191012,H191025, H191115,H191111,H191140,H191020,H191013,H191169,H191127,H191086,H191144,H191095,H191125,H191002,H191112,H191021,H191138,H191014,H191135,H191139,H191097,H191032,H191102,H191119,H191128,H191060,H191001,H191198,H191057,H191122,H191146,H191157,H191143,H191083,H191180,H191121,H191181,H191123,H191081,H191165,H191176,H191056,H191120,H191082,H191041,H191070,H191134,H191148,H191030,H191058,H191007,H191177,H191051,H191172,H191059 and H191037,  *** 12-12-12, AMENDED INFO ***:  H191167, H191171, H191184, H191185, H191192, H191194, H191202, H191205, H191207, H191208, H191211, H191214, H191217, H191219, H191220, H191222, H191223, H191224***

Root Cause

Device Design

Action Taken

Urgent Medical Device Correction/Urgent Field Safety Notices were sent on January 30, 2012 by certified mail to US and Canadian consignees. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** Varian Sales, Marketing and Service Organizations also were informed via letter the same day. The Field Safety Notice with customer distribution list was e-mailed to Varian International Customer Support Managers for translation (if necessary) to be sent to affected sites with return proof of notification to be sent to Varian. The notice informed users of the recommended actions. Questions were directed to the Varian Help Desk 1-888-827-4265 (US and Canada).