LiNA Loop Endoscopic Monopolar Loop for LSH (Laparoscopic Supracervical Hysterectomy). Large loop, 8mm connector, 1 X REF: EL-200-8 and 10 X REF: EL-200-8, Manufactured by LiNA Medical ApS, Surgery Division.

This recall has been terminated (originally issued June 28, 2011).

Company
Lina Medical Aps Formervangen 5 Glostrup Denmark
Recall Initiated
December 1, 2009
Posted
June 28, 2011
Terminated
February 23, 2012
Recall Number
Z-2690-2011
Quantity
1,950 pieces
Official Source
View on FDA website ↗

Reason for Recall

Potential for the LiNA Loop to break during operating procedure.

Distribution

GA, MN, PA, WA, TN, MA, IN, Ontario Canada

Lot / Code Info

Batch 9435, 9422, 9404, 9462, 9451, 9452, 9442, 9371, 9343, 9131, 9471, 7282, 8284, 8285, 8411, 8395, and 8442

Root Cause

Labeling design

Action Taken

Lina Medical Aps notified customers by newsletter in December 2009 followed by a letter to distributors including warnings added to instructions for use in January 2010.