Silicone CTS (Close to Shaft) Cuff Adjustable Neck Flange Tracheostomy Tube (Wire Reinforced), Arcadia Medical Silicone Air Cuff Adjustable Tracheostomy Tube, Sterile, Latex-Free, Provides direct tracheal access for airway management Arcadia Medical Corporation 1450 East American Lane, Schaumburg, IL 60173 USA Arcadia Medical Corporation Tortola, British Virgin Islands www.arcadiamedical.com, Made in USA
This recall has been terminated (originally issued June 24, 2011).
- Company
- Arcadia Medical Corporation
- Recall Initiated
- May 8, 2011
- Posted
- June 24, 2011
- Terminated
- August 30, 2011
- Recall Number
- Z-2657-2011
- Quantity
- 118
- Firm Location
- Crown Point, IN
- Official Source
- View on FDA website ↗
Reason for Recall
The product does not contain a punched hole above the cuff access port (a lumen that enables suctioning of secretions proximal to the inflated cuff) as intended by the devices design. This eliminates the user's ability to suction secretions through this suction port, necessitating that the user revert to standard of care methods for suctioning secretions from around the cuffed trach tube shaft.
Distribution
Worldwide Distribution --USA including the state of Ohio, and the countries of Denmark, Italy, Germany, Turkey, Slovakia, Isreal, France, Sweden, and Romania
Lot / Code Info
Lots 00911114 Catalog # 180-060, and Lot 00911116 Catalog # 180-080
Root Cause
Process control
Action Taken
Arcadia Medical Corporation sent a "MEDICAL DEVICE RECALL" letter on May 9, 2011 and email on May 8, 2011 to all affected costumers. The letter included; description of the product name, afffected lot numbers, description of the problem, instructions to discontinue use of the recalled devices and to return them to Arcadia Medical for replacement . The letter instructed the customers to email Arcadia Medical with the number of devices in their inventory. For additional information contact Arcadia Medical Corporation at 219-779-9431.