AC Power Cords, 110V Manufactured by Electri-Cord Manufacturing Company in Westfield, PA. Power Cord #4102204 used with Tricell 60601 Model 212857, First Step Select Model 215200 , and Plexipulse Model 205835.

This recall has been terminated (originally issued January 25, 2010).

Company
Kci Usa, Inc.
Recall Initiated
November 4, 2009
Posted
January 25, 2010
Terminated
January 21, 2011
Recall Number
Z-0663-2010
Quantity
10,520
Firm Location
San Antonio, TX
Official Source
View on FDA website ↗

Reason for Recall

Power cords may crack and fail inside plug with potential for fire hazard.

Distribution

Nationwide and Canada

Lot / Code Info

Device Serial Numbers Affected: Tricell 60601 - Serial Numbers FSUK00011-FSUK01400; used with First Step Select - Serial Numbers FWSK00001-FSWK02393; and used with Plexipulse - Serial Numbers PPGK00001-PPGK00226.

Root Cause

Nonconforming Material/Component

Action Taken

Kinetics Concepts, Inc. (KCI) issued an "Urgent Medical Device Correction" letter dated November 4, 2009. Consignees were informed of the affected product and instructed to take necessary action to avoid any further issues associated with the product. For further information, contact KCI's technical support team at 1-800-275-4524 option #3.