Glucose Oxidase Reagent as follows: Pointe Scientific Glucose Oxidase Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog # G7519-1L, G7519-500 and G7519-SAM. 2MM Diagnostics Glucose Oxidase Reagent Set, 2MM Corporation, Miami, FL; Reorder # RT-160-500. Pointe Scientific Glucose Oxidase Reagent, Pointe Scientific, Inc., Canton, MI; Catalog, MI; Catalog # 2-G7519-L, 7-G7519-100, 7-G7519-200, 7-G7519-1000, 8-G7519-100, and 12-RT-160-500. For the quantitative determination of Glucose in human serum. For in vitro diagnostic use only.

This recall has been terminated (originally issued March 12, 2009).

Company
Pointe Scientific, Inc.
Recall Initiated
November 3, 2008
Posted
March 12, 2009
Terminated
December 3, 2009
Recall Number
Z-1091-2009
Quantity
1500 liters
Firm Location
Canton, MI
Official Source
View on FDA website ↗

Reason for Recall

The product is unable to maintain the specification for linearity through the shelf life of the product.

Distribution

Worldwide Distribution -- USA, Hong Kong, Indonesia, Israel, Jordan, Mexico, Nigeria, Poland, Russia, Serbia, Sri Lanka, Trinidad, and Venezuela.

Lot / Code Info

All lots beginning in 722601, Exp. 08/09. (Lots 722601, 722601-009, 722601-017, 722601-035, 722601-057, 722601-072, 722601-081, 722601-084, 722601-085, 722601-092, 722601-093, 722601-094, 722601-134, 722601-136, 722601-168, 722601-185, 722601-232, 722601-233, 722601-254, 722601-255, 722601-318, 722601-337, and 722601-361.)

Root Cause

Process control

Action Taken

Consignees were notified via an Urgent: Medical Device Recall letter dated 11/3/08 that all test results obtained with these lots are questionable and that attending physicians should be informed to determine whether additional testing is required. Also, consignees should examine their inventories for any of the lot numbers listed and dispose of them. If the product has been further distributed, identify customers and notify them of the product recall by including a copy of the recall letter. The Recall Return Response Form should be completed and returned and upon receipt Pointe Scientific will replace any disposed of reagent. Questions should be directed to the Technical Support Department at 1-800-757-5313.