RecallDepth

Viasys Healthcare, Sonora Transcranial Doppler (TCD) System. Model Sonara. Part Number of Units: 982A0469 Sonara Unilateral. The Sonara Transcranial Doppler system is a medical ultrasound device for noninvasively measuring the blood flow velocities in arteries and veins.

This recall has been terminated (originally issued April 7, 2009).

Company
Cardinal Health, Inc
Recall Initiated
January 6, 2009
Posted
April 7, 2009
Terminated
July 3, 2012
Recall Number
Z-0998-2009
Quantity
36 systems
Firm Location
Madison, WI
Official Source
View on FDA website ↗

Reason for Recall

Internal testing has identified several several scenarios where unexpected System lockups, shutdowns or system errors would render the Sonara unavailable for use until the Sonara can be restarted. These issues can be corrected by installing the Version 6.0 b045 software updating the Audio Driver and updating to XP Service Pack 3. Additionally, improper wiring could cause the cooling fan to not op

Distribution

Worldwide Distribution -- USA including states of FL, DC, LA, NJ, NY, PA, TN, TX, WI, CA, MO, and LA and countries including THAILAND, SINGAPORE, SERBIA-MONTENEGRO, PORTUGAL, PHILIPPINES, POLAND, NORWAY, NETHERLANDS, KOREA, ITALY, ISRAEL, HUNGARY, HONG KONG, GREAT BRITAIN, SPAIN, GERMANY, CZECH REPUBLIC, SWITZERLAND, CANADA, BULGARIA, and BOSNIA-HERZEGOVINA.

Lot / Code Info

Serial Numbers:  PWH0297 PWH0212 PWH0221 PWH0309 PWH0246 PWH0250 PWH0223 PWH0429 PWH0245 PWH0292 PWH0298 PWH0214 PWH0211 PWH0210 PWH0219 PWH0216 PWH0224 PWH0215 PWH0213 PWH0220 PWH0253 PWH0252 PWH0254 PWH0221 PWH0218 PWH0217 06-2506 06-2504 05-2506 PWH0251 06-2501 06-2500 06-2503 06-2502 05-2502 and 05-2500.

Root Cause

Device Design

Action Taken

Consignees were sent a Cardinal Health "Urgent: Medical Device Correction" letter dated January 12, 2009 beginning the week of February 2, 2009 and continuing into week of February 9, 2009. The letter was addressed to customers describing the problems and systems affected and solutions. Cardinal Health requested customers to complete the actions listed in the letter to install the Field Correction and return the completed "Verification Form" via mail, e-mail or fax to Cardinal Health-NeuroCare. Further information can be obtained by calling 1-800-356-0007 ext 2191 or fax to 1-877-244-3805.

More recalls by Cardinal Health, Inc

View all recalls by this company →