RecallDepth

Monitor suspensions used with Advantx Legacy, RFX Classical, SFX-90, Model # 46-240838G2, manufactured after March 8, 1985 for Advantx Legacy, RFX and SFX systems. General-purpose radiographic system component used to suspend the system's image monitor.

Company
Ge Healthcare
Recall Initiated
March 9, 2007
Posted
October 27, 2008
Terminated
August 24, 2012
Recall Number
Z-0200-2009
Quantity
21
Firm Location
Waukesha, WI

Reason for Recall

GE Healthcare has recently become aware of a potential risk of monitor dropping associated with the monitor suspension of the Advantx system that may impact patient safety. It has been reported that suspensions manufactured after March 8, 1995 did not have the required thread locking agent applied to the setscrew, allowing the setscrew to back out over time, which can lead to the monitor dropping

Distribution

Worldwide Distribution --- USA including states of AZ, CA, FL, GA, IA, MD, MO, NC, NY, OH, OK, TN, TX,WI, and LA, and country of Slovak Republic.

Lot / Code Info

00000486566WK3, 00000486586WK1, 00000486555WK6, 00000486573WK9, 00000486581WK2, 00000478555WK6, 00000486579WK6, 00000486588WK7, 00000486557WK2, 00000486560WK6, 00000486550WK7, 00000486556WK4, 00000486547WK3, 00000056366VP8, 00000486544WK0, 00000486541WK6, 00000486543WK2, 00000471291WK5, 00000486568WK9, and 00000486559WK8

Root Cause

Component change control

Action Taken

Consignees were sent a GE Healthcare "Product Safety Notification" letter dated September 13, 2007. The letter was addressed to Hospital Administrators/Managers of Radiology/Radiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Corrective Action and Contact Information. The letter recommends that customers move the monitor suspension to an open area in the room where viewing is still possible and make sure to minimize the movement of the monitor until a Qualified GE Healthcare Field Engineer has inspected the suspension. The letter also states that a Qualified GE Healthcare Field Engineer will schedule a field visit to inspect the monitor suspension. Customers should contact their sales or service rep is they have any questions.