UniCel DxC 600/600PRO/600i/800/800PRO Synchron Clinical Systems , Part Numbers: A20463-Software version 1.0, A27331-Software version 1.2, A29764-Software version 1.4

This recall has been terminated (originally issued August 26, 2006).

Company
Beckman Coulter Inc.
Recall Initiated
May 15, 2006
Posted
August 26, 2006
Terminated
April 8, 2013
Recall Number
Z-1440-06
Quantity
524
Firm Location
Brea, CA
Official Source
View on FDA website ↗

Reason for Recall

Incorrect Reagent Status-When a new cartridge of Infinity Lithium Reagent is loaded on to the UniCel DxC System, the Reagent Status screen will incorrectly show 'Days Left' as 21. The correct Reagent Status for a new Lithium reagent cartridge is 14 days. Lithium reagent used past 14 days may produce low quality control and/or patient results.

Distribution

Nationwide and Canada

Lot / Code Info

Part Numbers: A20463-Software version 1.0, A27331-Software version 1.2, A29764-Software version 1.4

Root Cause

Other

Action Taken

A Product Correction Action (PCA) letter was sent the week of May 15, 2006 to all Unicel DxC customers informing them not to leave any Lithium reagent cartridge onboard a UniCel DxC System past 14 days. They also were instructed to complete and return the enclosed response form. The Letter was sent by US Mail.