RecallDepth

MicroScan Lens Disposable Plastic Cap used with MicroScan Video Microscope, MicroVisionMedical brand

This recall has been terminated (originally issued August 23, 2006).

Company
Microvision Medical Holding B.V. Meibergdreef 45 Amsterdam Netherlands
Recall Initiated
July 21, 2006
Posted
August 23, 2006
Terminated
November 15, 2006
Recall Number
Z-1435-06
Quantity
540 caps
Official Source
View on FDA website ↗

Reason for Recall

Some lenses may have microscopic holes in the disposable caps representing a potential break in the sterile barrier

Distribution

World wide-The product was shipped to physicians in UT, IA, OR, MA, LA, CA. The product was also shipped to Germany, United Kingdom, and The Netherlands.

Lot / Code Info

Lot number 583546

Root Cause

Other

Action Taken

The recalling firm issued a recall letter dated 7/21/06 to their customers informing them of the problem and the need to return the product.