Kodak DirectView DR 7100 System, Catalog #133-6049, a digital radiology system. The major system components are: operator console; digital imaging capture device; tilting table with floating top; ceiling-mounted x-ray tube; x-ray generator timing and distribution unit.
- Company
- Eastman Kodak Co
- Recall Initiated
- February 20, 2004
- Posted
- July 20, 2004
- Terminated
- May 28, 2004
- Recall Number
- Z-0728-04
- Quantity
- 3 units
- Firm Location
- Rochester, NY
Reason for Recall
Defective component in overhead tube support balancer may cause the suspended load to fall.
Distribution
Amsterdam, China and Kuwait.
Lot / Code Info
Serial Numbers affected: 4193-5702, 4193-5704, 4193-5705, 4193-5706, 4193-5708
Root Cause
Other
Action Taken
2/20/2004: Field representatives notified the customer in China that the unit should not be used until the OTS Balancer component can be repaired/replaced.