Kodak DirectView DR 7100 System, Catalog #133-6049, a digital radiology system. The major system components are: operator console; digital imaging capture device; tilting table with floating top; ceiling-mounted x-ray tube; x-ray generator timing and distribution unit.

This recall has been terminated (originally issued July 20, 2004).

Company
Eastman Kodak Co
Recall Initiated
February 20, 2004
Posted
July 20, 2004
Terminated
May 28, 2004
Recall Number
Z-0728-04
Quantity
3 units
Firm Location
Rochester, NY
Official Source
View on FDA website ↗

Reason for Recall

Defective component in overhead tube support balancer may cause the suspended load to fall.

Distribution

Amsterdam, China and Kuwait.

Lot / Code Info

Serial Numbers affected: 4193-5702, 4193-5704, 4193-5705, 4193-5706, 4193-5708

Root Cause

Other

Action Taken

2/20/2004: Field representatives notified the customer in China that the unit should not be used until the OTS Balancer component can be repaired/replaced.