The product is a sheath meant to fit over a Nasopharyngoscope and is labeled as EndoScrub 2 Sheaths

Company
Medtronic Xomed, Inc.
Recall Initiated
January 29, 2004
Posted
July 20, 2004
Terminated
May 3, 2012
Recall Number
Z-0954-04
Quantity
7 boxes, 5 sheaths per box
Firm Location
Jacksonville, FL

Reason for Recall

The sheaths in question have shafts that are one inch longer than required for the endoscope specified on the label.

Distribution

Products were distributed to 23 hospitals and surgical centers in NC, NB, IN, UT, TX, NY, WI, MO, TN, ID, OK, FL, MA, WY, NJ, LA.

Lot / Code Info

Lots 31396400 and 30722300

Root Cause

Other

Action Taken

A Recall notification letter was mailed to each hospital account on January 29, 2004 for each type of device they received which was under recall explaining the error and asking the accounts to remove recalled product from inventory, fax them the response sheet and to return the products.