RecallDepth

iLab Ultrasound Imaging System, Models 120INS and 240INS, manufactured by Boston Scientific Corporation, Fremont, CA 94538

Company
Boston Scientific Corporation
Recall Initiated
April 11, 2000
Posted
September 23, 2008
Terminated
April 6, 2009
Recall Number
Z-1693-2008
Quantity
120 units
Firm Location
Fremont, CA

Reason for Recall

Improperly terminated wires on a component of the iLab Acquisition processor power supply could cause the processor to lock-up or stop, and prolong the patients procedure.

Distribution

Product was distributed to a total of 81 consignees in the USA, Hong Kong, Great Britain, Austria and France.

Lot / Code Info

UPNs: H74900036130, H749FG000940, H749ILAB120INS0, H749ILAB120INSR0, H749ILAB220INS0, H749240INS0, M004EPIL120INS0. Serial numbers 0000004082, 0000004385, 0000004386, 0000004387, 0000004388, 0000004389, 0000004390, 0000004391, 0000004392, 0000004393, 0000004396, 0000004397, 0000004398, 0000004399, 0000004416, 0000004417, 0000004418, 0000004419, 0000004420, 0000004421, 0000004422, 0000004423, 0000004424, 0000004425, 0000004426, 0000004427, 0000004428, 0000004429, 0000004430, 0000004431, 0000004432, 0000004433, 0000004434, 0000004435, 0000004436, 0000004439, 0000004441, 0000004442, 0000004443, 0000004444, 0000004445, 0000004446, 0000004447, 0000004448, 0000004453, 0000004454, 0000004455, 0000004456, 0000004468, 0000004469, 0000004472, 0000004473, 0000004480, 0000004487, 0000004488, 0000004489, 0000004490, 0000004491, 0000004494, 0000004495, 0000004496, 0000004497, 0000004498, 0000004505, 0000004506, 0000004507, 0000004512, 0000004513, 0000004514, 0000004515, 0000004516, 0000004520, 0000004521, 0000004522, 0000004523, 0000004524, 0000004525, 0000004526, 0000004529, 0000004538, 0000004539, 0000004548, 0000004549, 0000004550, 0000004551, 0000004556, 0000004557, 0000004558, 0000004559, 0000004564, 0000004565, 0000004566, 0000004567, 0000004572, 0000004573, 0000004574, 0000004575, 0000004580, 0000004581, 0000004582, 0000004583, 0000004584, 0000004585, 0000004586, 0000004587, 0000004610, 0000004702, 0000004703, 0000004704, 0000004705, 0000004717, 0000004718, 0000004719, 0000004720, 0000004726, 0000004728, 0000004729, 0000004730, 0000004731, 0000004732, 0000004737, 0000004738, 0000004751, 0000035228, 0000035229, 0000035390, 0000035391, 0000036314, 0000094718, 0009430893, 0009503458, 0009543381, 0009565345, 0009902558, 0009602559, 0009626567, 0009640091, 0009828054, 0009828055, 0009828344, 0011008866.

Root Cause

Process control

Action Taken

On April 4 and 10, 2008, customer notifcations were sent to consignees. Field corrections are estimated to begin on April 25, 2008.