Sunrise Medical Quickie Groove Power Wheelchair with Transit option The product is a powered wheelchair for human use.

This recall has been terminated (originally issued September 23, 2008).

Company
Sunrise Medical Inc
Recall Initiated
November 13, 2006
Posted
September 23, 2008
Terminated
December 22, 2009
Recall Number
Z-2339-2008
Quantity
36 units
Firm Location
Fresno, CA
Official Source
View on FDA website ↗

Reason for Recall

Rear bolts holding the frame in place may shear and the front bolts may potentially come off the posts, resulting in partial or full seat detachment in a sudden stop or crash.

Distribution

Worldwide: US, Canada, New Zealand and Australia.

Lot / Code Info

Multiple serial numbers, starting with GRV-100: 182, 181, 178, 176, 166, 161, 160, 151, 150, 115, 110, 109, 106, 100, 099, 098, 094, 082, 081, 079, 076, 073, 064, 058, 056, 054, 053, 050, 044, 043, 041, 040, 035, 033, 021 and 013

Root Cause

Device Design

Action Taken

Initial phone contacts began on November 13, 2006. Letters were also mailed. The letter informed the customer of the problem and provided interim steps to take, until an transit upgrade kit is installed. Response cards will be tracked to measure response rate and assure receipt of notice. Contact Sunrise Medical at 1-800-456-8168 for assistance.