MIM 4.0 and MIM 4.1 is a software package designed for use in diagnostic imaging. It is a stand alone software package which operates on Windows 2000/XP.
- Company
- Mimvista Corp
- Recall Initiated
- April 23, 2008
- Posted
- September 3, 2008
- Terminated
- April 14, 2015
- Recall Number
- Z-1827-2008
- Quantity
- 221
- Firm Location
- Cleveland, OH
Reason for Recall
In MIM (4.0.2-4.0.5) and MIM (4.1.0-4.1.2) versions, there is an error that may occur during the calculation of statistics using the SUV Tool or Contour Statistics. It will occur only under the following specific circumstance: A fusion of a primary volume which has a Z-Axis slice separation larger than that of the secondary volume and statistics are reported on the secondary volume.
Distribution
The product was sold in the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI & WV. It was also sold in Europe, Asia and Canada.
Lot / Code Info
Model #'s 4.0.2-4.0.5 & 4.1.0-4.1.2
Root Cause
Software change control
Action Taken
MIMvista Corp has issued this safety recall notification for MIM versions 4.0.2 through 4.1.2. An error has been discovered in the way SW and contour statistics are measured and reported under some specific circumstances. Users of these versions should immediately contact MIMvista support at (866) 421-2536 or visit httv://www.mimvista.com/recall/ to obtain complete instructions regarding this error. A new version that corrects this problem is available on the Downloads page, http://www,mimvista.com/download