Electrode Adaptor Cable Quick-Combo System, Model 9055 The 9055 Electrode Adaptor is intended for use by trained personnel who use a Cardiac Science Automated External Defibrillator (AED) for defibrillation with Medtronic QUICK-COMBO Electrodes.

This recall has been terminated (originally issued September 17, 2008).

Company
Cardiac Science Corporation
Recall Initiated
January 23, 2008
Posted
September 17, 2008
Terminated
December 8, 2011
Recall Number
Z-1714-2008
Quantity
63
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

Instructions For Use is not clear about how the AED voice and text prompts will differ when the adapter is used.

Distribution

Distributed Nationwide US and Canada and Internationally to: Norway and Guam

Lot / Code Info

Lot numbers: 123004, 92804, 62305, 63005, 52504, 50405, 33105, 12605, 12505, 7715, 6952, and 6838

Root Cause

Labeling design

Action Taken

On 1/23/08 the firm issued a letter titled "9055 Electrode Adaptor IFU Update" to consignees sent via certified return receipt requested. Enclosed with the letter is an updated Instructions For Use, part number 70-00180-01, revision C. Consignees were advised to discard their old Instructions For Use. For assistance, contact Cardiac Science at 1-888-466-8686.