Spacelabs Medical Temperature Probe Adaptor PN-700-0031-00. Device is packaged in clear plastic bags with an attached insert card Insert card reads "TruLink series Dual Temperature Cable REF 700-0032-00 Spacelabs Medical Issaquah, WA USA"

This recall has been terminated (originally issued September 20, 2008).

Company
Spacelabs Healthcare, Incorporated
Recall Initiated
July 2, 2008
Posted
September 20, 2008
Terminated
June 7, 2011
Recall Number
Z-2314-2008
Quantity
3,733
Firm Location
Issaquah, WA
Official Source
View on FDA website ↗

Reason for Recall

Intermittent contact of temperature probe adaptors can cause a loss of temperature monitoring which can cause a delay in care when an alarm is missed.

Distribution

Nationwide and Worldwide to medical facilities.

Lot / Code Info

Lot numbers: OSC 01/07, OSC 02/07, OSC 03/07, OSC 04/07 and OSC 07/07.

Root Cause

Nonconforming Material/Component

Action Taken

Starting July 3, 2008 the firm issued a "MEDICAL DEVICE CORRECTION" letter notifying their customers of the potential hazards with the use of probe adapter, PN 700-031-00. All letters have a return receipt to verify each customer received the medical device correction letter. The firm will replace all PN 700-0031-00 adaptors with a replacement adaptor from their initial distributor. The firm initiated sending new adaptors separately, by mail, with a copy of the same Medical Device Correction letter to the firm's customers. The firm has requested their customers to throw away any adapter that is being recalled. Technical Support will be available for customer questions and technical questions and to ensure each unit is upgraded. Contact Spacelabs Healthcare at 425-657-7200 ext 5970 if you have questions.