Puritan-Bennett 800 Series Ventilator Backup Power Source (BPS) Model 802, Puritan-Bennett Corporation, A subsidiary of Mallinckrodt Inc., Carlsbad, CA

This recall has been terminated (originally issued September 16, 2008).

Company
Covidien Limited
Recall Initiated
March 21, 2008
Posted
September 16, 2008
Terminated
October 29, 2008
Recall Number
Z-1692-2008
Quantity
6 Ventilators and 151 factory replacement units
Firm Location
Boulder, CO
Official Source
View on FDA website ↗

Reason for Recall

Wiring in battery back-up power supply may short and cause thermal damage to ventilator.

Distribution

Nationwide, including one VA facility in IL and two military facilities in NC and NY. No foreign distribution.

Lot / Code Info

4-070523-SP Battery Assembly 4-840120DIUU-US 4-070520-SP BPS

Root Cause

Process control

Action Taken

All customers were notified by Federal Express letter and were contacted by phone through field service engineers. They were told to examine their inventory for any installed or spare part battery assemblies and to remove and quarantine these parts. Any ventilators with installed affected battery assemblies were to be taken out of service.