Philips OmniDiagnost Eleva X-Ray System is a diagnostic imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications.

This recall has been terminated (originally issued September 16, 2008).

Company
Philips Medical Systems North America Co. Phillips
Recall Initiated
March 22, 2008
Posted
September 16, 2008
Terminated
September 17, 2010
Recall Number
Z-1374-2008
Quantity
22 units
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

Unexpected movement of the X-ray table may occur. These movements may lead to patient injury, e.g., an uncontrolled tilt may cause the patient to slip off.

Distribution

Nationwide Distribution -- including states of CA, CO, KY, TN, LA, MI, FL, MS, TX, VA, WI, MO, AZ, and NY.

Lot / Code Info

Product number 9896 002 04432. SN number: 43, 49, 52, 107, 123, 131, 152, 157, 171, 179, 185, 202, 266, 333, 337, 389, 408, 446, 485, 491, 514, 515; Site ID #: 506241, 500046, 521648, 522220, 520020, 533112, 533314, 533471, 533530, 535257, 535118, 535924, 540860, 541490, 539106, 547017, 547142, 549798, 547348, 550258, 553676, and 553235.

Root Cause

Software design

Action Taken

On 3/22/08, the firm mailed letters to their customers via certified mail. The letter informed the customers of the recall, advising the customers to, "Use the emergency stop button or release the initiated movement command to halt undesired table movement. If you perform digital and conventional acquisition, you can avoid the problem by closing the single examinations daily or by clearing the patient database regularly." The letter further informed the customers that a Philips representative will be installing updated software at a future date. Contact Sarah Baxter at 425-487-7665 or Philips Call Center at 800-722-9377, #5, #2 if you have questions.