Cook Hilal Embolization Microcoil, 6 mm, sterile. The device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral arteriovenous malformations.

This recall has been terminated (originally issued September 16, 2008).

Company
Cook, Inc.
Recall Initiated
May 8, 2008
Posted
September 16, 2008
Terminated
October 21, 2008
Recall Number
Z-2058-2008
Quantity
180
Firm Location
Bloomington, IN
Official Source
View on FDA website ↗

Reason for Recall

The product length is declared incorrectly in mm when the unit of measurement should be cm.

Distribution

Nationwide, Australia, Brazil, Canada, Columbia, Hong Kong, India, Israel, Korea, Malaysia, Panama, Peru, Singapore, Taiwan, Uruguay and Venezuela.

Lot / Code Info

Lots F2100451, F2102327, F2102038, F2102833, F2109137, F2112851, F21114679, F2122527, F2125284, F2126416, F2136920, F2141567, F2142403, F2142578, F2143038, F2146823, F2146882, F2147969, F2150692, F2159335, F2159263, F2161387, F2172446, F2188499, F2189353, F2203609, F2204861, F2207480, F2213136, F2213468 and F2215748..

Root Cause

Labeling Change Control

Action Taken

Consignees were notified via recall letter dated 5/9/08 that the firm would replace product on hand. Distributors were requested to notify their customers of the recall. If you have questions, contact Customer Relations Department at 800-346-2686.