da Vinci S Surgical System, Model number IS2000 Endoscopic Instrument Control System, manufactured by Intuitive Surgical, Sunnyvale, CA.

This recall has been terminated (originally issued September 16, 2008).

Company
Intuitive Surgical, Inc.
Recall Initiated
June 1, 2008
Posted
September 16, 2008
Terminated
July 29, 2009
Recall Number
Z-2204-2008
Quantity
112 systems
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Defective software chip may cause the system to fail and lock up.

Distribution

Worldwide Distribution --- including USA and countries of Belgium, Bulgaria, Canada, China, Denmark, France, Germany, Italy, Korea, Romania, Russia, Saudi Arabia, Singapore, Spain, Sweden, and United Kingdom.

Lot / Code Info

Serial numbers SG099, SG145, SG162, SG205, SG294, SG295, SG298, SG299, SG300, SG301, SG302, SG303, SG304, SG305, SG306, SG307, SG308, SG309, SG310, SG311, SG312, SG313, SG314, SG315, SG316, SG317, SG318, SG319, SG320, SG321, SG322, SG323, SG324, SG325, SG326, SG327, SG328, SG329, SG330, SG331, SG332, SG333, SG334, SG35, SG336, SG337, SG39, SG340, SG341, SG342, SG323, SG344, SG345, SG346, SG347, SG348, SG349, SG350, SG351, SG352, SG353, SG355, SG356, SG357, SG358, SG359, SG360, SG361, SG362, SG363, SG364, SG365, SG367, SG368, SG369, SG370, SG371, SG373, SG374, SG375, SG376, SG379, SG380, SG381, SG382, SG383, SG384, SG385, SG386, SG387, SG388, SG390, SG393, SG394, SG395, SG396, SG397, SG399, SG400, SG401, SG402, SG403, SG404, SG405, SG406, SG407, SG408, SG409, SG410, SG411, SG412, and SG413

Root Cause

Software design (manufacturing process)

Action Taken

Recall initiated June 1, 2008. Urgent Device Recall Letters were sent via e-mail, with follow up phone calls and service visits to correct the product. The firm recommended that continued use of the da Vinci S Surgical System during the time period prior to our service update to replace the software chip. The letter emphasizes the importance of always having backup equipment and instrumentation available when performing surgery with the da Vinci S Surgical System and being prepared to convert to alternative surgical techniques, if necessary. Customers are asked to fill out and return the acknowledgement form.